uDL1 and uDL2 miniature data loggers

uDL1 and uDL2 miniature data loggers for cathodic protection monitoring. Features, measurement ranges, battery operation and USB data export.

Measurement and logging

CorTalk loggers record readings at cathodic protection test stations. Values can be viewed on the LCD or downloaded via USB. Both units measure 89 × 40 × 18 mm, weigh approximately 75 g and operate from −20 to 60 °C.

The lithium-ion battery can power the logger for up to approximately six months per charge or until the memory is full. GPS availability depends on the model and version. The receiver synchronises the clock and records the coordinates of the test location.

CorTalk uDL1 data logger with display

uDL1 potential measurements

This single-channel unit measures DC and AC voltage. Advanced G1 and G2 support GPS synchronisation and fast DC logging at up to 10 Hz. Basic stores up to 500,000 readings, G1 up to one million and G2 up to two million.

The logger starts automatically when its leads are connected. AC voltage measurements can be recorded at up to 1 Hz. The LCD shows readings, battery status and GPS connection status.

CorTalk uDL1

uDL2 voltage and current measurements

Two channels allow the operator to monitor coupon potential and current together. G1 provides ±60 and ±300 mA DC current ranges. G2/G3 provide three ranges: ±7, ±70 and ±300 mA.

The device has two galvanically isolated measurement channels with a shared common. It records both current and voltage from a cathodic protection coupon.

CorTalk uDL2

Data logger features

  • Long battery life, up to six months of continuous operation between charges.
  • Surge protection and protection against lightning-induced transients.
  • Compact cases that fit inside most test stations.
  • LCD for viewing readings without connecting a computer.
  • USB configuration and data download.
  • GPS time synchronisation and geographical coordinates for each measurement, on GPS-equipped versions.
  • Operating temperature range from −20 °C to +60 °C for field use.
  • Moisture resistance.
Data logger comparison
ParameteruDL1uDL2
VersionsBasic
Advanced G1
Advanced G2
G1
G2/G3
Memory500 000 (Basic)
1 000 000 (Advanced G1)
2 000 000 (Advanced G2)
2 000 000
GPSYes
(only for versions Advanced G1 and G2)
Yes
Recording interval1 s to 12 h1 s to 12 h
Fast sampling (10 readings per second)Yes
(only for versions Advanced G1 and G2)
No
Current measurementNoYes
Measurement channels12
Voltage measurement ranges3 ranges:
Low: ±150 mV DC / 110 mV AC (True RMS)
Med: ±20 V DC / 14 V (True RMS)
High: ±150 V DC / 106 VAC (True RMS)
1 range:
±22 V DC, 15 V RMS AC
Current measurement ranges-Model G1
Medium: +/-60mADC, 40mA true RMS AC,
High: +/-300mADC, 200mA true RMS AC,
Model G2/G3
Low: +/-7mADC, 5mA true RMS AC,
Medium: +/-70mADC, 50mA true RMS AC,
High: +/-300mADC, 200mA true RMS AC,
Measurement resolution1µV DC and AC1µV/1µA DC and AC
DC voltage accuracyLow: +/-0.25% + 100uV
Med: +/-0.5% + 12mV
high (-75V to +75V): +/-0.5% + 100mV
high (<-75V and >+75V): +/-1% + 100mV
+/-0.55% + 5mV
AC voltage accuracyLow +/-0.3% + 80uV, min.180uV Med: +/-0.65% + 10mV, min. 22mV
High (-75V to +75V): +/-0.65% + 80mV, min. 170mV
High (<-75V and >+75V): +/-1% + 80mV, min. 170mV
+/-1% + 5mV, min. 20mV
DC current accuracy-Low: +/-0.5% + 2uA
Med.: +/-0.5% + 20uA
High: +/-0.5% + 750uA
AC current accuracy-Low: +/-1% + 20uA, min. 50uA
Med.: +/-1% + 25uA, min. 100uA
High:+/-1% + 750uA, min. 750uA
Input impedance>20 MΩ>20 MΩ
BatteryLi-IonLi-Ion
Operating timeUp to approximately 6 months on one charge, or until memory is fullUp to approximately 6 months on one charge, or until memory is full
Dimensions89 × 40 × 18 mm89 × 40 × 18 mm
WeightApprox. 75 gApprox. 75 g